Operation 33
If a semi-automatic tare (net) is entered, it is not possible to access
the enter preset tare function.
If a preset tare is entered, it is still possible to access the
semiautomatic tare (net) function. The two different types of tare
are added.
All the semi-automatic tare (net) and preset tare functions will be
lost when the instrument is turned off.
3.3 Semi-Automatic Zero (Weight Zero-setting For Small
Variations)
By closing the SEMI-AUTOMATIC ZERO input, the weight is set to zero. The
zero setting will be lost when the instrument is turned off.
This function is only allowed if the weight is lower than the 0 set value (see 0
SET in Section 2.3 “Zero Parameters” on page 20), otherwise the t
-----
alarm
appears and the weight is not set to zero.
3.4 Peak
By keeping the input closed the maximum weight value reached remains
displayed. Opening the input the current weight is displayed.
If you wish to use this input to view a sudden variation peak, set
the FILTER ON THE WEIGHT (see Section 2.2) to 0.
3.5 Alarms
Display Description
ErCEL Load cell is not connected or is incorrectly connected; the load cell signal
exceeds 39 mV; the conversion electronics (A/D converter) is
malfunctioning.
Er OL Weight display exceeds 110% of the full scale.
Er Ad Internal instrument converter failure; check load cell connections, if
necessary contact Technical Assistance.
---------
Weight exceeds the maximum weight by 9 divisions.
Er OF Maximum displayable value exceeded (value higher than 999999 or lower
than -999999).
t
-------
Weight too high: zero setting not possible.
NAH-PU This message appears in the test weight setting, in weight (span)
calibration, after the fifth test weight value has been entered.
Err0r The value set for the parameter is beyond the permitted values; press
to quit the setting mode leaving the previous value unchanged.
Examples:
-a number of decimals is selected for full scale which exceeds the
instrument's display potential;
-value above the maximum setting value;
-the weight value set in test weight verification does not match the
detected mV increase.
bLOC Lock active on menu item, keypad or display.
nOdl SP It’s not possible to display properly the number because is greater than
999999 or less than -999999.
Table 3-1. Alarm Descriptions
Note
Note
0
ESC