3.6 Removing the Battery
Use the following procedure to remove the battery from the weigher. See Figure 4 for component locations.
1. Remove the 6 screws (70) around the outer edge of the battery compartment cover (69).
2. Lift out battery (66) and unplug connector.
3.7 Inspection and Cleaning
All weigher parts should be carefully cleaned and inspected prior to assembly to assure proper working
conditions.
All parts, with the exception of all electronic components, should be cleaned with kerosene only. Do not use
carbon tetrachloride, acetone, or any other organic cleaning agent. The use of organic-based agents is ex-
tremely detrimental to the nonmetallic elements within the unit and will cause premature failure. Electronic
components should be removed from the weigher and cleaned in alcohol and thoroughly dried before reas-
sembly. While cleaning electronic components, do not disturb the factory settings of the potentiometers.
Under no circumstances should the weigher be cleaned with steam or pressure type cleaners. Avoid using
caustic or acid base solutions as these can damage seals and other components.
3.8 Inspection Prior to Assembly
Use the following procedure to prior to reassembling the weigher platform. See Figure 5 on page 16 for
location of numbered parts.
1. Inspect piston pivot (13) for excessive wear and proper fitting.
2. Inspect platform buttons (43) for excessive wear.
3. Inspect platform load strips (18) and (19) for signs of unusual wear at contact points and proper fitting
of resilient retainer (17). Depth of wear must not exceed 0.002 inch.
Figure 4 — Battery Removal
66
69
70